1.
Introduction
|
c.
Decelerometers
|
a.
Exhaust Gas Analysers
|
| The MOT
Testing Conditions of Appointment stipulate the minimum test
bay equipment and other requirements. The MOT list of Acceptable
Equipment lists the various makes and models of test equipment
that have a measuring capability eg. headlamp aim testers, brake
testers etc. The suitability of other test equipment is individually
assessed at each testing station. |
Must
be calibrated at intervals not exceeding 2 years. This can be
done by a decelerometer manufacturer or a National Measurement
Accreditation Service (NAMAS) |
A
technician accredited by NAMAS. There are two exceptions |
| |
i.
non-catalyst OIML Class I equipment with an approved self-gassing
facility requires a 6 monthly NAMAS calibration with an immediate
calibration which may be performed by testing station staff. |
| |
ii.
equipment (non-catalyst or catalyst) with an approved automatic
monthly self-gassing facility requires a 12 monthly NAMAS calibration
with a monthly calibrations which may be performed by testing
station staff. |
2.
Equipment Maintenance and calibration
|
d.
Tyre Tread Depth Gauges
|
b.
Roller and Plate Brake Testers and Headlamp Aim Testers
|
| All MOT
testing equipment must be properly maintained and kept in good
order. Measuring equipment must also be calibrated and shown
to be accurate at the appropriate time intervals (see paragraph
3) and in all cases calibration records must be kept, except
for tyre tread depth gauges. |
A
periodic accuracy check is required. This can be done by placing
the base foot onto a flat surface (eg. window glass and checking
that the zero datums align. |
Either |
| |
i.
the original manufactureror |
| |
ii.
a firm which has taken over responsibility for equipment support
from the manufacturer,or |
| |
iii.
a calibration specialistor |
| |
iv.
an experienced person from the testing station. The Vehicle
Inspectorate will witness a calibration prior to confirmation
of this arrangement and may subsequently require periodic demonstration
of continuing competence. |
3.
Frequency and Location of Calibration
|
e.
Diesel smoke meters
|
5.
Documentation
|
a.
Exhaust Gas Analysers
|
i.
Daily (checks by the user)
Carry out a leak check on the hose and probe. Most machines
have an automatic facility for doing this. If not, a physical
check of the hose and probe must be carried out to confirm that
there are no leaks.
ii. Three, six or twelve monthly - depending on make/type
An exhaust gas analyser must be calibrated with a certified
calibration gas and when the ambient temperature is above 5
deg C (see item 4a)
Note: Gas analysers cannot be properly calibrated if the mains
AC voltage is less than 204v or more than 264v. |
i.
Weekly checks by the user
ii. Annual calibration by a technician accredited by NAMAS. |
In
cases where a testing station employs either the manufacture
or a calibration specialist, there must be a written contract
and the test station must obtain a certificate of each calibration |
b.
Roller Brake Testers and Headlamp Aim Testers
|
4.
Calibration Personnel
|
|
| Must
be calibrated in situ every 6 months. It is acceptable for calibration
certificates to be issued at any time during every sixth month. |
Equipment
may be calibrated on site only by the following personnel: |
|